TOP TOP

Gradient Shimming With 1H Signal

Most autoshimming with gradients (gradient shimming) is done with a strong deuteron (2H) signal from deuterated solvent. In cases where a strong 2H signal isn't present when the solvent isn't deuterated, a strong 1H signal can be used to perform gradient-aided shimming. This is often the case for biological samples dissolved in water, either pure H2O or ~90% H2O/10% D2O. If a deuteron signal isn't present at all, lock has to be turned off; the experiment must be run un-locked. General background on gradient shimming can be found here. The following outlines the steps to perform gradient shimming with a strong 1H signal.

NOTE:

Steps:

1. Collect a simple 1H spectrum.

Assuming the presence of a strong 1H signal, receiver gain has to be dropped, likely all the way to zero and the flip angle has to be reduced drastically to prevent receiver and ADC overflow. You may enter gain=0 pw=1 nt=1 go. Save the 1H spectrum for referencing purpose later.

2. Enter gmapsys. Go to Acquire->Gradient Shim

Alternatively, generate your own shim map with current sample

You may want to generate your own shim map with the current sample and shim using this map later. To generate a map:

3. Quit gmapsys and gradient shim experment

  • After shimming, click Quit Gradient Autoshim. The experiment should return to the original experiment.

  • Updated, July 2019, H. Zhou

    ©2019 The Regents of the University of California. All Rights Reserved. UC Santa Barbara, Santa Barbara, CA 93106