Lithium Detection

Lithium has two isotopes, Li6 and Li7, with a natural abundance of 7.6% and 92.4%, respectively. Li6 (spin=1) has a 50-fold smaller quadrupolar moment than Li7 (spin=3/2), making Li6 signal sharper than Li7, particularly in an organometallic system with bound lithium. But Li6 sensitivity is 420 fold lower than Li7 at natural abundance, making both Li6 and Li7 useful depending on the resolution and sensitivity tradeoff. Li6 enrichment should yield both better linewidth and sensitivity. Both nuclei have a chemical shift range from -16 to 11ppm.

Referencing (for both Li6 and Li7)

As usual, choose one of several referencing methods:

  1. For rough referencing with lock turned on, type setref
  2. Add TMS (or DSS for aqueous solution), reference TMS/DSS 1H signal to 0ppm (in exp1). In Li6/7 experiment (in exp2), type mref(1,2) or use other methods as discuussed here.
  3. Use a separate sample (9.7M LiCl in D2O) to reference Li6/7 peak to 0ppm. Without shimming, then collect data from your sample under same experiment, running unlocked.

One-pulse Li7 detection on NMR500 with SWP probe

Li7 experiment should be run locked if possible. If you have done Li6 experiment with the lock cable disconnected, re-connect it. Ideally, you want to have have two samples:

  1. The actual sample with your compound
  2. a 9.7M LiCl sample in D2O for testing and referencing

Step 1: Lock and shim

Step 2: Tune probe and replace 1/4-wavelength cable

Step 3: Collect data

Step 4: Finishing off: reconnect cables and tune X-channel back to C13

One-pulse Li6 detection on NMR500 with SWP probe

Note that the resonance frequency of Li6 is very close to that of deuterium with a ratio of 1:1.043. With the filters available on nmr500, it is recommended that Li6 experiment is run unlocked as instructed below.

Ideally, you want to have two samples:

  1. The actual sample with your compound
  2. A Li6/7 sample with 9.7M LiCl in D2O. This is used for referencing and testing if needed.

Step 1: Lock and shim

Step 2: Tune probe, disconnect lock cable and replace 1/4-wavelength cable

Step 3:Collect data

Step 4: Finishing off: reconnect cables and tune X-channel back to C13


H. Zhou updated Dec 2010